User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Fluke 5200A Sub 3 DATE: 24-Oct-97 AUTHOR: User Contributed REVISION: ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 18 NUMBER OF LINES: 139 CONFIGURATION: HP 5335A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 HEAD {* Frequency Tests *} 1.002 JMP 2.001 1.003 EVAL 2.001 DISP Connect the UUT HI/LO outputs to the 5335A 2.001 DISP CH A input. Do not connect the UUT sense lines. 2.002 DISP Set the UUT SENSE switch to INT. 2.003 IEEE [TERM LF] 2.004 IEEE [@5335]FN1;AU1; 2.005 DISP Change the UUT as follows: 2.005 DISP VOLTAGE RANGE ...................... 1V 2.005 DISP VOLTAGE dials ...................... 1.000000 2.005 DISP FREQUENCY RANGE .................... 100 Hz 2.005 DISP FREQUENCY dials .................... 100.00 2.006 ACC 100.000H 1e-10U 2.007 IEEE [@5335][I] 2.008 MEME 2.009 MEMC H 1.2U #! T.U.R. not calculated because System Uncertainty too small. 3.001 DISP Change the UUT as follows: 3.001 DISP FREQUENCY RANGE .................... 1 KHz 3.002 ACC 1.0000kH 1e-10U 3.003 IEEE [@5335][I] 3.004 MEM/ 1000 3.005 MEME 3.006 MEMC H 0.012U #! T.U.R. not calculated because System Uncertainty too small. 4.001 DISP Change the UUT as follows: 4.001 DISP FREQUENCY RANGE .................... 10 KHz 4.002 ACC 10.000kH 1e-10U 4.003 IEEE [@5335][I] 4.004 MEM/ 1000 4.005 MEME 4.006 MEMC H 0.120U #! T.U.R. not calculated because System Uncertainty too small. 5.001 DISP Change the UUT as follows: 5.001 DISP FREQUENCY RANGE .................... 100 KHz 5.002 ACC 100.00kH 1e-10U 5.003 IEEE [@5335][I] 5.004 MEM/ 1000 5.005 MEME 5.006 MEMC H 1.2U #! T.U.R. not calculated because System Uncertainty too small. 6.001 DISP Change the UUT as follows: 6.001 DISP FREQUENCY RANGE .................... 1 MHz 6.002 ACC 1.0000MH 1e-10U 6.003 IEEE [@5335][I] 6.004 MEM/ 1000000 6.005 MEME 6.006 MEMC H 0.035U #! T.U.R. not calculated because System Uncertainty too small. 7.001 HEAD {* Frequency Linearity Test *} 7.002 JMP 8.001 7.003 EVAL 8.001 DISP Change the UUT as follows: 8.001 DISP FREQUENCY RANGE .................... 1 KHz 8.001 DISP FREQUENCY dials .................... 1.1111 8.002 ACC 1.1111kH 1e-10U 8.003 IEEE [@5335][I] 8.004 MEM/ 1000 8.005 MEME 8.006 MEMC H 0.013U #! T.U.R. not calculated because System Uncertainty too small. 9.001 DISP Change the UUT as follows: 9.001 DISP FREQUENCY dials .................... 0.9999 9.002 ACC 999.90H 1e-10U 9.003 IEEE [@5335][I] 9.004 MEME 9.005 MEMC H 12U #! T.U.R. not calculated because System Uncertainty too small. 10.001 DISP Change the UUT as follows: 10.001 DISP FREQUENCY dials .................... 0.8888 10.002 ACC 888.80H 1e-10U 10.003 IEEE [@5335][I] 10.004 MEME 10.005 MEMC H 11U #! T.U.R. not calculated because System Uncertainty too small. 11.001 DISP Change the UUT as follows: 11.001 DISP FREQUENCY dials .................... 0.7777 11.002 ACC 777.70H 1e-10U 11.003 IEEE [@5335][I] 11.004 MEME 11.005 MEMC H 10U #! T.U.R. not calculated because System Uncertainty too small. 12.001 DISP Change the UUT as follows: 12.001 DISP FREQUENCY dials .................... 0.6666 12.002 ACC 666.60H 1e-10U 12.003 IEEE [@5335][I] 12.004 MEME 12.005 MEMC H 9U #! T.U.R. not calculated because System Uncertainty too small. 13.001 DISP Change the UUT as follows: 13.001 DISP FREQUENCY dials .................... 0.5555 13.002 ACC 555.50H 1e-10U 13.003 IEEE [@5335][I] 13.004 MEME 13.005 MEMC H 8U #! T.U.R. not calculated because System Uncertainty too small. 14.001 DISP Change the UUT as follows: 14.001 DISP FREQUENCY dials .................... 0.4444 14.002 ACC 444.40H 1e-10U 14.003 IEEE [@5335][I] 14.004 MEME 14.005 MEMC H 7U #! T.U.R. not calculated because System Uncertainty too small. 15.001 DISP Change the UUT as follows: 15.001 DISP FREQUENCY dials .................... 0.3333 15.002 ACC 333.30H 1e-10U 15.003 IEEE [@5335][I] 15.004 MEME 15.005 MEMC H 6U #! T.U.R. not calculated because System Uncertainty too small. 16.001 DISP Change the UUT as follows: 16.001 DISP FREQUENCY dials .................... 0.2222 16.002 ACC 222.20H 1e-10U 16.003 IEEE [@5335][I] 16.004 MEME 16.005 MEMC H 5U #! T.U.R. not calculated because System Uncertainty too small. 17.001 DISP Change the UUT as follows: 17.001 DISP FREQUENCY dials .................... 0.1111 17.002 ACC 111.10H 1e-10U 17.003 IEEE [@5335][I] 17.004 MEME 17.005 MEMC H 4U #! T.U.R. not calculated because System Uncertainty too small. 18.001 DISP Set the UUT MODE switch to STBY. 18.002 DISP Disconnect the UUT from the counter.